Remote control commands
R&S
®
RTP
1666User Manual 1337.9952.02 ─ 12
Return values:
<Count> Total number of decoded frames.
Usage: Query only
BUS<m>:LIN:FRAMe<n>:STATus?
Returns the overall state of the selected frame.
Suffix:
<m>
.
1..4
Selects the serial bus.
<n> *
Selects the frame.
Return values:
<FrameState> OK | UART | CHCKsum | VERS | LENer | SPERror | PRERror |
SYERror | WAKeup | CPERror | INSufficient | INComplete
UART: at least one UART error occured. LIN uses UART words
without parity bit.
CHCKsum: checksum error
VERS: the version of the LIN standard is not valid
LENer: unexpected length
SPERror: stop error
PRERror: parity error in identifier
SYERror: synchronization error
WAKeup: the frame is a wakeup frame
CPERror: parity error and checksum error
INSufficient: the frame is not completely contained in the acqui-
sition. The decoded part of the frame is valid.
INComplete: the frame is missing some parts
*RST:
OK
Usage: Query only
BUS<m>:LIN:FRAMe<n>:STARt?
BUS<m>:LIN:FRAMe<n>:STOP?
Returns the start time and stop time of the selected frame, respectively.
Suffix:
<m>
.
1..4
Selects the serial bus.
<n> *
Selects the frame.
Return values:
<Start>, <Stop> Range: -100E+24 to 100E+24
*RST: 0
Default unit: s
Protocols