Remote control commands
R&S
®
RTP
1813User Manual 1337.9952.02 ─ 12
BUS<m>:RFFE:COUPling <ThresCpl>
If enabled, the SDATA thresholds are set to the SCLK threshold values.
Suffix:
<m>
.
1..4
Parameters:
<ThresCpl> ON | OFF
*RST: ON
BUS<m>:RFFE:DATA:THReshold:HIGH <ThresDatPos>
Set the positive going threshold (V
TP
) for the data signal.
Note that the high and low thresholds and the hysteresis values are dependent, so it is
sufficient to define two values for each line.
Suffix:
<m>
.
1..4
Parameters:
<ThresDatPos> Range: 0 to 2.5
Increment: 0.1
*RST: 0.72
Default unit: V
BUS<m>:RFFE:DATA:THReshold:LOW <ThresDatNeg>
Set the negative going threshold (V
TN
) for the data signal.
Note that the high and low thresholds and the hysteresis values are dependent, so it is
sufficient to define two values for each line.
Suffix:
<m>
.
1..4
Parameters:
<ThresDatNeg> Range: 0 to 2.5
Increment: 0.1
*RST: 0.48
Default unit: V
BUS<m>:RFFE:DATA:THReshold:HYSTeresis <ThresDatHyst>
Set the hysteresises between V
TP
and V
TN
for the data signal.
Note that the high and low thresholds and the hysteresis values are dependent, so it is
sufficient to define two values for each line.
Suffix:
<m>
.
1..4
Protocols