Remote control commands
R&S
®
RTP
1844User Manual 1337.9952.02 ─ 12
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<ProbeCP> DIFFerential | SINGle
*RST: SINGle
BUS<m>:DPHY:CP:SOURce <ClockSource>
Selects the source of the clock lane.
Usually, the source is one of the analog channels. Reference and math waveforms are
available as source if the trigger source is one of the analog channels but not the serial
bus.
Alternatively, digital channels can be used if MSO option R&S RTP-B1 is installed. Dig-
ital and analog channels cannot be used at the same time.
For triggering on a serial bus, analog or digital channel sources are required.
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<ClockSource> NONE | C1W1 | C2W1 | C3W1 | C4W1 | M1 | M2 | M3 | M4 |
M5 | M6 | M7 | M8 | R1 | R2 | R3 | R4
*RST: NONE
BUS<m>:DPHY:DRATe <DataRate>
Sets a data rate.
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<DataRate> Range: 1000000 to 2500000000
Increment: 1000000
*RST: 800000000
Default unit: bps
BUS<m>:DPHY:DSPData <DispDatVect>
Enables the display of the data vector, which means that the whole frame will be deco-
ded.
When the display vector is disabled only the packet header will be decoded and not the
actual data payload of the frame. This will speed up the decoding.