Remote control commands
R&S
®
RTP
1858User Manual 1337.9952.02 ─ 12
Usage: Query only
BUS<m>:DPHY:RESult:FRAMe<n>:PACKet<o>:VALue?
Returns the byte value of the data payload.
Suffix:
<m>
.
1..4
Selects the serial bus.
<n> *
Selects the frame.
<o> *
Selects the packet number.
Return values:
<PacketValue> Range: 0 to 9223372036854775808
Increment: 1
*RST: 0
Usage: Query only
BUS<m>:DPHY:RESult:FRAMe<n>:STARt?
Returns the start time of the specified frame.
Suffix:
<m>
.
1..4
Selects the serial bus.
<n> *
Selects the frame.
Return values:
<Start> Range: -100E+24 to 100E+24
Increment: 100E-12
*RST: 0
Default unit: s
Usage: Query only
BUS<m>:DPHY:RESult:FRAMe<n>:STATe?
Returns the overall state of the frame.