Remote control commands
R&S
®
RTP
1888User Manual 1337.9952.02 ─ 12
ONE
Sets the trigger condition to binary 1.
ZERO
Sets the trigger condition to binary 0.
DC
Sets the trigger condition to any CREQ value (don't care).
*RST: DC
TRIGger<m>:MPHY:RREQ <REQ>
Specifies the RREQ bit pattern.
Suffix:
<m>
.
1..3
Parameters:
<REQ> ONE | ZERO | DC
ONE
Sets the trigger condition to binary 1.
ZERO
Sets the trigger condition to binary 0.
DC
Sets the trigger condition to any RREQ value (don't care).
*RST: DC
TRIGger<m>:MPHY:CREDit:CONDition <CreditOperator>
Sets the condition for Credit. You can define an exact value or a value range.
Suffix:
<m>
.
1..3
Parameters:
<CreditOperator> EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less or equal than, Greater Than,
Greater or equal than. These conditions require one data pattern
to be set with the corresponding TRIGger<m>:MPHY:CREDit:
MIN command.
INRange | OORange
In range/Out of range: set the minimum and maximum value of
the range with TRIGger<m>:MPHY:CREDit:MIN and
TRIGger<m>:MPHY:CREDit:MAX.
*RST:
EQUal
Protocols