EasyManua.ls Logo

Rohde & Schwarz RTP Series - Page 1962

Rohde & Schwarz RTP Series
2403 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTP
1962User Manual 1337.9952.02 ─ 12
Return values:
<Start> Range: -100E+24 to 100E+24
Increment: 100E-12
*RST: 0
Default unit: s
Usage: Query only
BUS<m>:CMSB:RESult<n>:CELL<o>:STOP?
Returns the end time of the selected cell.
Suffix:
<m>
.
1..4
Selects the serial bus.
<n> *
<o> *
Selects the cell.
Return values:
<Stop> Range: -100E+24 to 100E+24
Increment: 100E-12
*RST: 0
Default unit: s
Usage: Query only
BUS<m>:CMSB:RESult<n>:CELL<o>:STATe?
Returns the state of the cell.
Suffix:
<m>
.
1..4
<n> *
<o> *
Return values:
<State> OK | LENGth | UNKNown | INComplete | CRC | PARity
OK
No error detected
LENGth
The length error indicates that the cell ended prematurely. The
amount of bits that the software expected (based upon the
user's frame description) was not found before the cell was ter-
minated. This might occur because a new frame synchronized,
or a gap appeared between the bits.
UNKNown
Unknown error
Protocols

Table of Contents

Other manuals for Rohde & Schwarz RTP Series

Related product manuals