Remote control commands
R&S
®
RTP
2186User Manual 1337.9952.02 ─ 12
Parameters:
<TimeRefSlope> POSitive | NEGative | EITHer
*RST:  EITHer
Usage:  Asynchronous command
EYE<m>:TREFerence:LEVel <TimeRefLevel>
Sets the reference level on the timing reference waveform to define the timestamps for
slicing the data waveform. A timestamp is set each time, when the selected slope of
the timing waveform crosses the selected reference level.
Suffix: 
<m>
.
1..4
Parameters:
<TimeRefLevel> UPPer | MIDDle | LOWer
*RST:  MIDDle
Usage:  Asynchronous command
EYE<m>:HORizontal[:UNIT] <TimeUnits>
Sets the method to define the horizontal scale of the eye diagram.
Suffix: 
<m>
.
1..4
Parameters:
<TimeUnits> ATIMe | UINTerval
ATIMe
Absolute time
UINTerval
Unit intervals. You can use unit interval, if the nominal frequency
of the clock signal is known.
*RST:  UINTerval
Usage:  Asynchronous command
EYE<m>:HORizontal:ABSolute:RANGe <HorizontalRange>
Sets the time range that is covered by the eye diagram.
Suffix: 
<m>
.
1..4
Parameters:
<HorizontalRange> Range:  100E-12  to  100
Increment:  1E-9
*RST:  200E-9
Default unit: s
Protocols