Remote control commands
R&S
®
RTP
2253User Manual 1337.9952.02 ─ 12
Parameters:
<IQCarrierFreq> Range: 10E-3 to 20E+9
Increment: 100
*RST: 1E+6
Default unit: Hz
Usage: Asynchronous command
CHANnel<m>:IQ:SBRF <IQSidebandRF>
Defines the frequency position of the RF spectrum in the input signal: normal or
inverse. The position is important for correct down-conversion and filtering.
Prerequisite: CHANnel<m>:IQ:INPType on page 2250 is set to REAL
For details, see "Sideband (real input)" on page 1008
Suffix:
<m>
.
1..4
Selects the input channel.
Parameters:
<IQSidebandRF> INVerse | NORMal
*RST: NORMal
Usage: Asynchronous command
CHANnel<m>:IQ:SBIF <IQSidebandIF>
Defines the sideband and the frequency position of complex modulated input signal in
IF range.
Prerequisites: CHANnel<m>:IQ:INPType on page 2250 is set to COMPlex and
CHANnel<m>:IQ:INPMode on page 2251 is set to RFIF
For details, see "Sideband (complex IF input)" on page 1008
Suffix:
<m>
.
1..4
Selects the input channel.
Parameters:
<IQSidebandIF> NUPPer | NLOWer | IUPPer | ILOWer
NUPPer: Upper sideband in normal position
NLOWer: Lower sideband in normal position
IUPPer: Upper sideband in inverse position
ILOWer: Lower sideband in inverse position
*RST: NUPPer
Usage: Asynchronous command