Remote control commands
R&S
®
RTP
2273User Manual 1337.9952.02 ─ 12
ADVJitter<m>:RESult:UNIT <UnitType>
Selects between the absolute and relative unit for the displayed results.
Suffix:
<m>
.
1..2
Parameters:
<UnitType> ABS | REL
*RST: ABS
ADVJitter<m>:RESult:COMPonents<n>:STATistics:MAXPeakpeak? <PeakValue>
ADVJitter<m>:RESult:COMPonents<n>:STATistics:MAXStddev? <StdDevValue>
ADVJitter<m>:RESult:COMPonents<n>:STATistics[:MINStddev]? <StdDevValue>
ADVJitter<m>:RESult:COMPonents<n>:STATistics:MINPeakpeak? <PeakValue>
ADVJitter<m>:RESult:COMPonents<n>:STATistics:PEAKpeak?
ADVJitter<m>:RESult:COMPonents<n>:STATistics:WFMCount? <CountValue>
ADVJitter<m>:RESult:COMPonents<n>:STATistics:STDDev?
Return the statistic results of the specified measurement.
MAXPeakpeak: absolute maximum value of the component.
MAXStddev: standard deviation of the maximum value.
MINPeakpeak: absolute minimum value of the component.
MINStddev: standard deviation of the minimum value.
PEAKpeak: peak to peak value of the long-term measurement results.
WFMCount: number of waveforms (acquisitions) the measurement is based on.
STDDev: standard deviation of the long-term measurement results.
Suffix:
<m>
.
1..2
<n> 1..29
See Table 22-20
Return values:
<StdDev> Range: -100E+24 to 100E+24
Increment: 100E-12
*RST: 0
Usage: Query only
ADVJitter<m>:RESult:PERiodic<n>:COUNt?
Returns the number of periodic components.
Suffix:
<m>
.
1..2
<n> *
Not relevant