Waveform generator (Option R&S RTP-B6)
R&S
®
RTP
970User Manual 1337.9952.02 ─ 12
Content and format of the R&S arbitrary generator *.bin files
The file stream should contain the following information in the given order:
●
Sample rate
[double format]
●
Number of samples
[unsigned integer format]
●
Samples
[double format] * number of samples
15.1.5.1 General settings
Waveform source
Selects the arbitrary waveform source. You can load an existing file or load the current
oscilloscope waveform.
Remote command:
WGENerator<m>:ARBGen[:SOURce] on page 2232
Run mode
Selects the duration for which the signal of the arbitrary generator will be output after
the trigger event. You can choose between a "Continuous" and "Single period" dura-
tion.
Remote command:
WGENerator<m>:ARBGen:RUNMode on page 2231
Setup of the waveform generator