7 Basic Functions for Scan Head Control and Laser Control
172
Testing 3-Axis Scan Systems with F-Theta
Objective
• With the adjusted 3D correction table and
3D vector commands, see Chapter
”3D Commands”, page 238, mark two identical
large squares. Mark one of the squares with the
work piece in z position z = z
min
, the other one
with z = z
max
.
• These two squares should match exactly. If this is
not the case, your correction file is not perfectly
suited to your objective. To solve this problem,
measure the size (x and y) of both squares and
report these values to SCANLAB. You will then
receive a new 3D correction file.
• See also Section ”Enhanced 3D Correction”,
page 240.