INTRODUCTION, LCD Screen
Sysmex SF-3000 Operator's Manual -- Revised September 1995 1-11
NOTE: To set the next Sample ID No., Rack No. and Tube Position No. to be
analyzed in the auto mode by means of Sampler Unit, press [Sampler]
keypad to display the auto mode setting window.
The DP No. is displayed below the system status. If the Data Printer (DP) is the selected
output device, the DP No. displayed indicates the next sample data number to be printed
out.
Errors -- "2.0 kg/cm
2
Error", "Rack Move Error", etc. is displayed below the system
status, in place of the DP No. If 2 or more errors occur simultaneously, they are
displayed in order of importance. When an error is displayed, press [Sysmex] key to
display the action/help messages provided in the Error Log program. Refer to Chapter 8:
Troubleshooting for the action message display.
• Data Processing Status
X M : Indicates the status of X M control (one file of the QC program) in the
following way:
No indication : X M control is not in use (Stop).
X M :XM control is in use (Start).
X M (with red back-light) : X M Limit Error has occurred.
HC : Indicates transmission status of data to the Host Computer.
No indication : The Host Computer is set to "Not Connected".
HC : Ready for transmission to Host Computer.
HC (with green back-light) : Now transmitting to Host Computer.
HC (with red back-light) : Transmission to Host Computer not occurring
because of abnormality in Host Computer.
HC (with yellow back-light) : Error ("Not Transmissible", "No Ack" or "Ack
Code Err") has occurred between SF-3000 and
Host Computer.
GP : Indicates the status of the Graphic Printer (or Line Printer).
No indication : The Graphic Printer is set to "Not Connected".
GP : The Graphic Printer is ready.
GP (with green back-light) : Data being output to Graphic Printer.
GP (with red back-light) : Printing not possible because of abnormality in
Graphic Printer.
DP : Indicates the status of the Data Printer.
No indication : The Data Printer is set to "Not Connected".
DP : The Data Printer is ready.
DP (with green back-light) : Data being output to Data Printer.
DP (with red back-light) : Printing not possible because of abnormality in
Data Printer.