TROUBLESHOOTING
8-50 Sysmex SF-3000 Operator's Manual -- Revised September 1995
HC Communication Error
Phenomena A communication problem exists between the Main Unit and the host
computer.
Probable Cause 1) The power of the host computer is OFF.
2) The host computer is disconnected or improperly connected.
3) The host computer is not ready for data transmission.
Corrective Action 1) Check the host computer power and the system status.
2) Verify the cable between the Main Unit and the host computer is securely
connected.
3) Set the host computer connection to "Not Connected", then reset to
"Connected". Refer to
Chapter 11, Section 5.1: HC Setting
.
4) If the error persists, contact Sysmex service representative.
Not Assigned
Phenomena The output to HC is assigned when the host computer is not specified for
connection.
Output to HC is impossible until the error is resolved.
Probable Cause HC has not been specified in Settings program.
Corrective Action 1) Set the HC to "Connected". Refer to
Chapter 11, Section 5.1: HC Setting
for the procedures.
2) If the error persists, contact Sysmex service representative.