TROUBLESHOOTING
Sysmex SF-3000 Operator's Manual -- Revised September 1995 8-51
Not Transmissible
Phenomena The output data is not received by the host computer.
Probable Cause 1) The HC setting is not correct.
2) The HC has a problem or is down (not functioning).
3) The interface cable is disconnected or improperly connected.
Corrective Action 1) Check the HC settings. Refer to
Chapter 11, Section 5.1: HC Setting
.
2) Check the HC status in the Status Display by pressing [Status] key in the
root menu.
3) Check that the interface cable is securely connected.
4) If the error persists, contact Sysmex service representative.
Off Line
Phenomena The output to HC is assigned when the DSR signal from the host computer is
OFF.
Output to HC is impossible until the error is resolved.
Probable Cause 1) The host computer power is OFF.
2) The HC is in the off line condition.
3) The interface cable is disconnected or improperly connected.
Corrective Action 1) Check that the host computer is powered ON and is on-line.
2) Check that the interface cable is securely connected.
3) If the error persists, contact Sysmex service representative.