EasyManuals Logo

SYSMEX SF-3000 User Manual

SYSMEX SF-3000
605 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #393 background imageLoading...
Page #393 background image
TROUBLESHOOTING
8-72 Sysmex SF-3000 Operator's Manual -- Revised September 1995
(4) Press the Start Switch. Rack shift operation will begin. Check the sampler
operation.
,,,

Rack Shift
Measurement Li
Analysis Line
Figure 8-15: Rack Shift Operation
(5) Press [Return] to end the test program.
3 . Rack Feed Out
Select this program to execute the Rack Feed Out operation test.
(1) Set the rack in place in the extreme left of the sampler analysis line.
(2) Bring the Test screen by pressing [Mainte.] and then [Test] keypads (Refer to
Section 3.1 in this chapter).
(3) Press the [Rack Feed Out] keypad from the Test screen. The message "Set Rack
and Press Start Switch" will appear.
(4) Press the Start Switch. Rack feed out operation will begin. Check the sampler
operation.
,,,

Rack Feed out
Figure 8-16: Rack Feed Out Operation
(5) Press [Return] to end the test program.