TROUBLESHOOTING
8-74 Sysmex SF-3000 Operator's Manual -- Revised September 1995
3.7 Test Bar Code Reader
Use this program to test the ID Read function.
(1) Place bar-coded tubes in a sampler rack, and place the rack in the right rack pool.
(2) Bring the Test screen by pressing [Mainte.] and then [Test] keypads (Refer to
Section 3.1 in this chapter).
(3) Press the [ID Read] keypad from the Test screen. The message "Set Rack and
Press Start Switch" will appear.
NOTE: If the ID reader is set to "Not Used" in the settings program, the [ID
Read] key is masked and cannot be accessed.
(4) Press the Start Switch.
Rack feed in, rack shift, and then ID read operations will be executed, and the
Sample No. read will be displayed on the LCD. Verify that the sample No. is read
correctly.
Sysmex
DP:1234567890123
XM HC GP DP SU
Ready
Manual Mode Next No.1234567890123
Sampler
Return
Maintenance
RACK TUBE SAMPLE NO.
ID Read
1
1
1
1
1
1
1
1
1
1
1
2
3
4
5
6
7
8
9
10
1234567890123
1234567890124
1234567890125
1234567890126
1234567890127
1234567890128
1234567890129
1234567890130
1234567890131
1234567890132
Figure 8-18: ID Read Display
(5) Press [Return] to end the test program.