TROUBLESHOOTING
Sysmex SF-3000 Operator's Manual -- Revised September 1995 8-75
3.8 Test Memory
This program provides a display of the stored data check window, and shows the test
results.
(1) Bring the Test screen by pressing [Mainte.] and then [Test] keypads (Refer to
Section 3.1 in this chapter).
(2) Press the [Memory] keypad from the Test screen. The stored data check window
will appear showing the check results of the Stored Data, Stored Data Table,
Setting Data and QC Data.
Sysmex
DP:1234567890123
XM HC GP DP SU
Ready Test Mode Next No.1234567890123
Sampler
Maintenance
< Test >
SRV
WB Motor
Sheath Motor
Rinse Motor
Mixing Motor
PDA
Rack Feed In
Rack Shift
Rack Feed Out
Sampler CP Mech.
ID Read
Memory
Floppy Disk
LCD Display
Touch Panel
DP Print
GP Print
HOST Interface
Laser
Return
< Memory Test >
OK
STORED DATA
STORED DATA TABLE
SETTING DATA
QC DATA
OK
OK
OK
OK
Figure 8-19: Stored Data Check Window
"OK" indicates that data is normal. "NG" indicates that there is an abnormality in
the data. If "NG" is displayed, contact your Sysmex service representative.
(3) Press [OK] key to end the Memory Test program.
(4) Press [Return] to end the Test program.