TROUBLESHOOTING
Sysmex SF-3000 Operator's Manual -- Revised September 1995 8-77
(5) If the floppy disk is abnormal, an error message will appear.
Sysmex
DP:1234567890123
XM HC GP DP SU
Ready Test Mode Next No.1234567890123
Sampler
Maintenance
< Test >
SRV
WB Motor
Sheath Motor
Rinse Motor
Mixing Motor
PDA
Rack Feed In
Rack Shift
Rack Feed Out
Sampler CP Mech.
ID Read
Memory
Floppy Disk
LCD Display
Touch Panel
DP Print
GP Print
HOST Interface
Laser
Return
< Floppy Disk Test >
OK
FD format NG.
Figure 8-21: When FD Test Error Occurred
NOTE: A full floppy disk or unformatted floppy disk is assumed as abnormal
and the message "FD format NG" will be displayed.
(6) Press [OK] to end the FD Test program.
(7) Press [Return] to end the Test program.