INSTALLATION, Manual CP Unit
B-20 Sysmex SF-3000 Operator's Manual -- Revised May 1996
(6) Connect the tubes and cables to
the Main Unit (see Figure
B-27) except the whole blood
aspiration tube.
(7) Fasten the fluid connector and
make sure to attach the rubber
seal and align protrusions.
Wiring Cord No.177
Figure B-27: Connect Tubes
CAUTION: • When assembling the sampler, connect the ground cable (green/yellow
cable) first.
• When disassembling the sampler, disconnect the ground cable last.
(8) Push the Manual CP unit all the way toward the Main Unit and make sure that the
MCP Unit is fastened securely with the catchers.
(9) Fasten the Manual CP Unit to the Main Unit Chassis with the screw binding (see
Figure B-28).
M3x6
Screw Binding
(M3 x 6)
Figure B-28: Install Manual CP Unit