SAMPLE PROCESSING, Capillary Mode Procedures
Sysmex SF-3000 Operator's Manual -- Revised September 1995 2-27
After the sample is aspirated, the analysis begins. You can check the status of each stage
of analysis using the Analysis Process screen. To see this screen, press [Status] key in
the root menu. Switch to "P2. Analysis Process" using the [
] and [
] keys.
Sysmex
DP:1234567890123
XM HC GP DP SU
Ready
Capillary Mode Next No.1234567890123
Sampler
STATUS
< Analysis Process >
Sequence
ID No.
Rack
Aspirate
Dilute
RBC Count
WBC/BASO Count
DIFF Count
RBC Analyze
WBC/BASO Analyze
DIFF Analyze
Edit
Store
13
123-567-90123
0000-00
12
123-567-90122
0000-00
11
123-567-90121
0000-00
10
123-567-90120
0000-00
: Entry
: Interruption
: Run
: Finish Abnormal
: Finish Normal
Return
Figure 2-17: Analysis Process Screen (Capillary Mode)
6 . Display Analysis Results
After analysis is completed (approximately 85 seconds after the sample is aspirated), the
results are displayed on the screen.
NOTE: "C" in green backlight is added to the left of Sample ID No.
XM
12345
Last
NO.
R B C P L T
D I F F W B C / B A S O
[g/dL]
[fL]
WBC
NEUT
LYMPH
MONO
EO
BASO
RBC
HGB
HCT
MCV
MCH
MCHC
RDW-SD
RDW-CV
PLT
PDW
MPV
P-LCR
250fL 40fL
RBC Flag
PLT Flag
WBC Flag
DP:1234567890123
Sysmex
Ready
Sampler
3
[x10 /uL]
[%]
[%]
[%]
[%]
[%]
6
[x10 /uL]
[g/dL]
[%]
[%]
[%]
[fL]
[fL]
[fL]
3
[x10 /uL]
[pg]
Capillary Mode
Next No.1234567890123
HC GP D P SU
7.23
4.62
1.55
0.71
0.18
0.17
4.74
14.1
41.6
87.8
29.7
33.9
46.4
15.7
213
9.8
10.3
24.3
63.9
21.4
9.8
2.5
2.4
R:0000-00
15:48 1996/6/30
Status
Work
List
QC
Stored
Data
GP
Print
Return
DP
Print
Replace
Reagent
C