EasyManuals Logo

Keysight Technologies M8000 Series User Manual

Keysight Technologies M8000 Series
668 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #146 background imageLoading...
Page #146 background image
146 Keysight M8000 Series of BER Test Solutions User Guide
4 User Interface - M8020A Display Views
Jitter
Sweep
profile, Step
distance
M1.DataOut1 Jitter sweep
mode
User programmable only If Mode is
selected as Variable amplitude
":SOURce:JITTer:SWEep:STATe
:SOURce:JITTer:SWEep:AMPLitude:MODE
:SOURce:JITTer:SWEep:STEP:DISTance
:SOURce:JITTer:SWEep:DATA:FILE"
Jitter
Sweep Step
Distance
M1.DataOut1 Jitter sweep
mode
Jitter sweep step distance
parameter is re-programmable only
if Mode is selected as Variable
amplitude
":SOURce:JITTer:SWEep:STATe
:SOURce:JITTer:SWEep:AMPLitude:MODE
:SOURce:JITTer:SWEep:STEP:DISTance"
CMI
Amplitude
M1.DataOut1 CMI State Value doesn't applicable if state is
Off
":SOURce:VOLTage:AMPLitude
:SOURce:INTerference:LEVel:CMODe:STATe
:SOURce:INTerference:LEVel:CMODe:AMPLitude"
Amplifier
Amplitude
The max value of CMI amplitude
depends on amplifier amplitude
":SOURce:VOLTage:AMPLitude
:SOURce:INTerference:LEVel:CMODe:STATe
:SOURce:INTerference:LEVel:CMODe:AMPLitude"
Amplifier Offset Depends on DataOut amplifier
offset, high and low values
":SOURce:VOLTage:AMPLitude
:SOURce:INTerference:LEVel:CMODe:STATe
:SOURce:INTerference:LEVel:CMODe:AMPLitude
:SOURce:VOLTage:LOW
:SOURce:VOLTage:HIGH
:SOURce:VOLTage:OFFSet"
DMI
Amplitude
M1.DataOut1 DMI State Value doesn't applicable if state is
Off
":SOURce:VOLTage:AMPLitude
:SOURce:INTerference:LEVel:DMODe:STATe
:SOURce:INTerference:LEVel:DMODe:AMPLitude"
Amplifier
Amplitude
The max value of DMI amplitude
depends on amplifier amplitude
":SOURce:VOLTage:AMPLitude
:SOURce:INTerference:LEVel:DMODe:STATe
:SOURce:INTerference:LEVel:DMODe:AMPLitude"
Amplifier Offset Depends on DataOut amplifier
offset, high and low values
":SOURce:VOLTage:AMPLitude
:SOURce:INTerference:LEVel:DMODe:STATe
:SOURce:INTerference:LEVel:DMODe:AMPLitude
:SOURce:VOLTage:LOW
:SOURce:VOLTage:HIGH
:SOURce:VOLTage:OFFSet"
Error
Insertion
Ratio
M1.DataOut1 Error Insertion
Mode
Error ratio of 1E-1 or 1E-2 are
selectable only if error insertion
mode is selected as Variable spacing
":OUTPut:EINSertion:MODE
:OUTPut:EINSertion:RATio"
Inter-Symb
ol
Interference
Mode
M1.DataOut1 Depends on
selected Preset
value
ISI mode as Two-Point only available
when Preset is selected as Custom
":SOURce:INTerference:ISYMbol:MODE
:SOURce:INTerference:ISYMbol:PRESet"
Functional
block /
Parameter
Identifier
Location
Dependent
Parameters
Description Related /Dependent SCPI

Table of Contents

Other manuals for Keysight Technologies M8000 Series

Questions and Answers: