146 Keysight M8000 Series of BER Test Solutions User Guide
4 User Interface - M8020A Display Views
Jitter
Sweep
profile, Step
distance
M1.DataOut1 Jitter sweep
mode
User programmable only If Mode is
selected as Variable amplitude
":SOURce:JITTer:SWEep:STATe
:SOURce:JITTer:SWEep:AMPLitude:MODE
:SOURce:JITTer:SWEep:STEP:DISTance
:SOURce:JITTer:SWEep:DATA:FILE"
Jitter
Sweep Step
Distance
M1.DataOut1 Jitter sweep
mode
Jitter sweep step distance
parameter is re-programmable only
if Mode is selected as Variable
amplitude
":SOURce:JITTer:SWEep:STATe
:SOURce:JITTer:SWEep:AMPLitude:MODE
:SOURce:JITTer:SWEep:STEP:DISTance"
CMI
Amplitude
M1.DataOut1 CMI State Value doesn't applicable if state is
Off
":SOURce:VOLTage:AMPLitude
:SOURce:INTerference:LEVel:CMODe:STATe
:SOURce:INTerference:LEVel:CMODe:AMPLitude"
Amplifier
Amplitude
The max value of CMI amplitude
depends on amplifier amplitude
":SOURce:VOLTage:AMPLitude
:SOURce:INTerference:LEVel:CMODe:STATe
:SOURce:INTerference:LEVel:CMODe:AMPLitude"
Amplifier Offset Depends on DataOut amplifier
offset, high and low values
":SOURce:VOLTage:AMPLitude
:SOURce:INTerference:LEVel:CMODe:STATe
:SOURce:INTerference:LEVel:CMODe:AMPLitude
:SOURce:VOLTage:LOW
:SOURce:VOLTage:HIGH
:SOURce:VOLTage:OFFSet"
DMI
Amplitude
M1.DataOut1 DMI State Value doesn't applicable if state is
Off
":SOURce:VOLTage:AMPLitude
:SOURce:INTerference:LEVel:DMODe:STATe
:SOURce:INTerference:LEVel:DMODe:AMPLitude"
Amplifier
Amplitude
The max value of DMI amplitude
depends on amplifier amplitude
":SOURce:VOLTage:AMPLitude
:SOURce:INTerference:LEVel:DMODe:STATe
:SOURce:INTerference:LEVel:DMODe:AMPLitude"
Amplifier Offset Depends on DataOut amplifier
offset, high and low values
":SOURce:VOLTage:AMPLitude
:SOURce:INTerference:LEVel:DMODe:STATe
:SOURce:INTerference:LEVel:DMODe:AMPLitude
:SOURce:VOLTage:LOW
:SOURce:VOLTage:HIGH
:SOURce:VOLTage:OFFSet"
Error
Insertion
Ratio
M1.DataOut1 Error Insertion
Mode
Error ratio of 1E-1 or 1E-2 are
selectable only if error insertion
mode is selected as Variable spacing
":OUTPut:EINSertion:MODE
:OUTPut:EINSertion:RATio"
Inter-Symb
ol
Interference
Mode
M1.DataOut1 Depends on
selected Preset
value
ISI mode as Two-Point only available
when Preset is selected as Custom
":SOURce:INTerference:ISYMbol:MODE
:SOURce:INTerference:ISYMbol:PRESet"
Functional
block /
Parameter
Identifier
Location
Dependent
Parameters
Description Related /Dependent SCPI