Keysight M8000 Series of BER Test Solutions User Guide 147
User Interface - M8020A Display Views 4
De-emphasi
s Preset
Register
Number
M1.DataOut1 State of Preset
Enable button
Re-programmable only if state of
Preset Enable button is ON
":OUTPut:DEEMphasis:PRESet:ENABle
:OUTPut:DEEMphasis:PRESet"
Bit
Recovery
Mode State
M1.DataIn1 Depends on CDR
State
BRM mode can only be enabled if
CDR state is ON
":INPut:CDR:JTF:STATe
:INPut:DATA:BRMode"
CDR State M1.DataIn1 Depends on CDR
Control type
If CDR control type is Sequence
controlled, then CDR state will be
controlled by Analyzer sequence
block settings
:INPut:CDR:CTRL
CDR Loop
Bandwidth
M1.DataIn1 Depends on Data
Rate
Maximum value of CDR loop
bandwidth depends on Data rate
(Synthesizer frequency)
":INPut:CDR:JTF:STATe
:SOURce:FREQuency"
Depends on Loop
Order
Depends on type of the selected
loop order
":INPut:CDR:JTF:STATe
:INPut:CDR:JTF:LORDer"
Depends on
Peaking
Peaking and Loop bandwidth are
interdependent
":INPut:CDR:JTF:LORDer
:INPut:CDR:JTF:SECond:PEAKing"
Comparator
Compare
Mode
M1.DataIn1 Depends on
Termination
Configuration
Available only if Termination
configuration is selected as
Unbalanced
:INPut:TCONfig
Termination
Voltage
M1.DataIn1 Depends on
Termination
Configuration
Available only if Termination
configuration is selected as
Unbalanced
":INPut:TCONfig
:INPut:TVOLtage
:INPut:CMODe"
SSC Profile
Shape
M1.ClkGen Depends on SSC
Profile
Available only if SSC profile is
selected as Arbitrary
":SOURce:SSCLocking:STATe
:SOURce:SSCLocking:PROFile
:SOURce:SSCLocking:SHAPe"
Functional
block /
Parameter
Identifier
Location
Dependent
Parameters
Description Related /Dependent SCPI

Table of Contents