148 Keysight M8000 Series of BER Test Solutions User Guide
4 User Interface - M8020A Display Views
Table 45 M8040A Dependent Parameters
Functional
block /
Parameter
Identifier
Location
Dependent
Parameters
Description Related /Dependent SCP
Clock Gen
Frequency
M1.ClkGen M8046A Clock
source
Depends on termination voltage,
CMI and DMI amplitudes
":SOURce:VOLTage:AMPLitude
:OUTPut:TVOLtage
:SOURce:INTerference:LEVel:CMODe:AMPLitude
:SOURce:INTerference:LEVel:DMODe:AMPLitude"
M8046A Follow
Sys Clock Option
Depends on amplitude and offset
value
":SOURce:VOLTage:HIGH
:SOURce:VOLTage:AMPLitude
:SOURce:VOLTage:OFFSet"
M1.ClkOut State Depends on amplitude and offset
value
":SOURce:VOLTage:LOW
:SOURce:VOLTage:AMPLitude
:SOURce:VOLTage:OFFSet"
M1.ClkOut
Divider
Depends on termination voltage ":SOURce:VOLTage:OFFSet
:OUTPut:TVOLtage "
Amplifier M1.DataOut1 Amplitude Depends on termination voltage and
synthesizer frequency
":SOURce:VOLTage:AMPLitude
:OUTPut:TVOLtage
:SOURce:FREQuency"
High Depends on amplitude and offset ":SOURce:VOLTage:HIGH
:SOURce:VOLTage:AMPLitude
:SOURce:VOLTage:OFFSet"
Low Depends on amplitude and offset ":SOURce:VOLTage:LOW
:SOURce:VOLTage:AMPLitude
:SOURce:VOLTage:OFFSet"
Offset Depends on termination voltage ":SOURce:VOLTage:OFFSet
:OUTPut:TVOLtage"
Coupling Selecting incorrect coupling enables
output protection, and as a result
DataOut state could not be enabled
":OUTPut:STATe
:OUTPut:COUPling"
Termination
Model
Only available when coupling is
selected as DC
":OUTPut:COUPling
:OUTPut:TCONfig"
Termination
Voltage
Available only if coupling is selected
as DC to termination model is
selected as unbalanced
":OUTPut:COUPling
:OUTPut:TCONfig
:OUTPut:TVOLtage"