Page #149 background imageLoading...
Page #149 background image
Keysight M8000 Series of BER Test Solutions User Guide 149
User Interface - M8020A Display Views 4
Clock Gen
Frequency
Maximum amplitude will be 900mV
if synthesizer frequency is below
32.414GHz, but if synthesizer
frequency is above 32.414GHz, then
maximum amplitude will be 600mV
":SOURce:VOLTage:AMPLitude
:SOURce:FREQuency"
De-emphasi
s
Co-efficient
2
M1.DataOut1 Automatic main
cursor state
Re-programmable only if Automatic
main cursor state is disabled
":OUTPut:DEEMphasis:CURSor:MAIN:AUTO
:OUTPut:DEEMphasis:CURSor:MAGNitude2"
De-emphasi
s
Co-efficient
s
M1.DataOut1 Depends on
Coefficient 0, 1
and 3 state
De-emphasis coefficient values are
interdependent and sum of all the
cursors cannot be more than 1.0
":OUTPut:DEEMphasis:CURSor:MAIN:AUTO
:OUTPut:DEEMphasis:CURSor:MAGNitude0
:OUTPut:DEEMphasis:CURSor:MAGNitude1
:OUTPut:DEEMphasis:CURSor:MAGNitude2
:OUTPut:DEEMphasis:CURSor:MAGNitude3"
LF Jitter M1.DataOut1 PJ State Value doesn't applicable if state is
Off
:SOURce:JITTer:LFRequency:PERiodic:STATe
PJ Amplitude PJ Amplitude depends on PJ
frequency
":SOURce:JITTer:LFRequency:PERiodic:STATe
:SOURce:JITTer:LFRequency:PERiodic:AMPLitude
:SOURce:JITTer:LFRequency:PERiodic:FREQuency"
PJ Frequency PJ frequency depends on PJ
amplitude
":SOURce:JITTer:LFRequency:PERiodic:STATe
:SOURce:JITTer:LFRequency:PERiodic:AMPLitude
:SOURce:JITTer:LFRequency:PERiodic:FREQuency"
Unit If unit is selected as sec, then
amplitude min/max values will be
expressed in sec
":SOURce:JITTer:LFRequency:UNIT
:SOURce:FREQuency"
HF Jitter M1.DataOut1 PJ1 State Value doesn't applicable if state is
Off
:SOURce:JITTer:HFRequency:PERiodic1:STATe
PJ1 Amplitude Depends on PJ2 amplitude, BUJ
amplitude and RJ amplitude
":SOURce:JITTer:HFRequency:PERiodic2:AMPLitude
:SOURce:JITTer:HFRequency:BUNCorrelate:AMPLitu
de
:SOURce:JITTer:HFRequency:RANDom:AMPLitude"
PJ2 State Value doesn't applicable if state is
Off
:SOURce:JITTer:HFRequency:PERiodic2:STATe
PJ2 Amplitude Depends on PJ1 amplitude, BUJ
amplitude and RJ amplitude
":SOURce:JITTer:HFRequency:PERiodic1:AMPLitude
:SOURce:JITTer:HFRequency:BUNCorrelate:AMPLitu
de
:SOURce:JITTer:HFRequency:RANDom:AMPLitude"
Functional
block /
Parameter
Identifier
Location
Dependent
Parameters
Description Related /Dependent SCP

Table of Contents

Other manuals for Keysight Technologies M8000 Series

Questions and Answers:

Question and Answer IconNeed help?