Power Meter Remote Operation 1
Keysight N8262A Programming Guide 103
How to Use Register
The polling method is used to access the information in status groups when:
– your language/development environment does not support SRQ interrupts.
– you want to write a simple, single purpose program and do not want to add the
complexity of setting an SRQ handler.
The Condition Polling Method
In this polling method, the power meter has a passive role. It only informs the
controller that conditions have changed when the controller asks. When you
monitor a condition with the polling method, you must:
1 Determine which register contains the bit that monitors the condition.
2 Send the unique query that reads that register.
3 Examine the bit to see if the condition has changed.
The polling method works well if you do not need to know about the changes the
moment they occur. The SRQ method is more effective if you must know
immediately when a condition changes. Detecting an immediate change in a
condition using the polling method requires your program to continuously read
the registers at very short intervals. This is not particularly efficient and there is a
possibility that an event may be missed.

Table of Contents

Other manuals for Keysight Technologies P Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies P Series and is the answer not in the manual?

Keysight Technologies P Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelP Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals