MEMory Subsystem 7
Keysight N8262A Programming Guide 309
MEMory:TABLe:GAIN[:MAGNitude]
<numeric_value>{,<numeric_value>}
This command is used to enter calibration factors into the sensor calibration table
(8480 Series sensors only) or offsets into the frequency dependent offset table,
currently selected using MEMory:TABLe:SELect. Any previous calibration factor
list, or offset list is cleared before the new calibration factors/offsets are stored.
A maximum of 81 parameters for sensor calibration tables and 80 parameters for
frequency dependent offset tables can be sent with this command. For sensor
calibration tables only, the first parameter is the reference calibration factor, each
subsequent parameter is a calibration factor point in the sensor calibration table.
Entries in the frequency lists correspond as shown in Table 7-4 with entries in the
calibration/offset factor lists.
For sensor calibration tables the number of frequency points must be one less
than the number of calibration factor data points. This is verified when the sensor
calibration table is selected using SENSe:CORRection:CSET1:SELect <string>.
Table 7-4 Frequency and Calibration/Offset factor list
Frequency Calibration Factor/Offset
-
Reference Calibration Factor
(For Sensor Calibration Tables)
Frequency 1 Calibration Factor/Offset 1
""
Frequency 80 Calibration Factor/Offset 80