SYSTem Subsystem 12
Keysight N8262A Programming Guide 489
Table 12-9 CDMAone: Power meter presets: Window/Measurement settings
Function Setting
Single Channel Dual Channel
Display setup
Upper window UU single numeric See Table 12-10
Lower window Dual numeric See Table 12-10
Window/measurement setup
Upper window/upper measurement (UU)
Feed Gate 1 Channel A
Gate 1 primary channel
[a]
Measurement Avg Avg
Upper window/lower measurement (UL)
Feed DEF See Table 12-10
Measurement DEF See Table 12-10
Lower window/upper measurement (LU)
Feed Gate 1 Channel A See Table 12-10
Measurement Peak See Table 12-10
Lower window/lower measurement (LL)
Feed Gate 1 Channel A See Table 12-10
Measurement Pk-to-Avg See Table 12-10
[a] For further information refer to “Primary and Secondary Channels” on page 473.