SENSe Subsystem 10
Keysight N8262A Programming Guide 349
Query
AVERage:COUNt? [MIN|MAX]
The query returns the current setting of the filter length or the values associated
with MIN and MAX. The format of the response is <NR1>.
Query Example
Error Messages
If a filter length value is entered using [SENSe[1]]|SENSe2:AVERage:COUNt while
[SENSe[1]]|SENSe2:SPEed is set to 200, the error –221, “Settings Conflict”
occurs. However, the filter length value is set but the
[SENSe[1]]|SENSe2:AVERage:STATe command is not automatically set ON.
AVER:COUN? This command queries the filter length for
Channel A.