SERVice Subsystem 16
Keysight N8262A Programming Guide 635
SERVice:SENSor[1]|2:PCALfactor <cal_factor_data>
This command writes calibration factor data to, or reads calibration factor data
from, the currently connected sensor. The whole calibration factor block must be
written at once as a checksum is generated. The new block must not be larger
than the existing block.
For E4410 Series and E9300 Series sensors, average path data, refer to
“SERVice:SENSor[1]|2:CDATe?” on page 631.
Syntax
Parameters
Query
SERVice:SENSor[1]|2:PCALfactor?
The query returns the current peak path calibration factor block.
SERV
1
2
:PCAL
:SENS
cal_factor_data
?
Space
Item Description/Default
cal_factor_data A binary data block. Refer to Appendix A, “Calibration Factor Block Layout,” on page A-673 for
further information.