14 TRIGger Subsystem
594 Keysight N8262A Programming Guide
TRIGger[:SEQuence]:LEVel:AUTO <boolean>
This command enables/disables automatic setting of the trigger level.
When this command is set to:
– ON, automatic setting of the trigger level is enabled.
– OFF, automatic setting of the trigger level is disabled.
– ONCE, automatic setting of the trigger level is enabled for one trigger event
only. The value is then set to OFF.
Syntax
Example
Reset Condition
On reset the value is set to ON.
TRIG:SEQ:LEV:AUTO 0 This command disables the automatic
setting of the trigger level for Channel A.
:LEV
0|OFF
1|ON
Space
?
:AUTOTRIG
:SEQ
ONCE