4 CALibration Subsystem
240 Keysight N8262A Programming Guide
Query
CALibration[1]|2:ZERO:NORMal:AUTO?
The query always returns a value of 0.
Error Messages
If zeroing was not carried out successfully the error –231, “Data Questionable;
ZERO ERROR” occurs. The error message will specify which channel failed
zeroing.
If this command is set to ON the error –224, “Illegal parameter value” occurs.
If there is no sensor connected, or if a sensor other than P-Series is connected,
the error –241, “Hardware missing” occurs.