MEMory Subsystem 7
Keysight N8262A Programming Guide 285
MEMory:CATalog[:ALL]?
This command lists stored sensor calibration tables (8480 Series sensors only),
frequency dependent offset tables and save/recall registers.
The power meter returns the data in the form of two numeric parameters and as
many strings as there are stored tables and save/recall registers:
<numeric_value>,<numeric_value>{,<string>}
The first numeric parameter indicates the amount of memory, in bytes, used
for the storage of tables and registers.
The second numeric parameter indicates the memory, in bytes, available for
the storage of tables and registers.
Each string parameter returned indicates the name, type and size of a stored
table or save/recall register:
<string>, <type>, <size>
<string> indicates the name of the table or save/recall register.
<type> indicates TABL for sensor calibration and frequency dependent
offset tables, or STAT for a save/recall register.
<size> indicates the size of the table or save/recall register in bytes.
A sample of a response may look like the following:
1178,26230,"DEFAULT,TABL,14","8481A,TABL,116",
"8482A,TABL,74",..........."State0,STAT,1619",
"State1,STAT,1619","State2,STAT,1619" ...........
The power meter is shipped with a set of predefined sensor calibration tables. The
data in these sensor calibration tables is based on statistical averages for a range
of Keysight Technologies power sensors. These tables can be edited. The
predefined data is listed in your User’s Guide. These power sensors and table
numbers are listed in Table 7-1.