14 TRIGger Subsystem
600 Keysight N8262A Programming Guide
Example
Reset Condition
On reset, TRIGger:DELay:AUTO is set to ON.
Query
TRIGger:DELay:AUTO?
The query enters a 1 or 0 into the output buffer indicating the status of
TRIGger:DELay:AUTO.
– 1 is returned when it is ON
– 0 is returned when it is OFF
Query Example
TRIG:SEQ:DEL:AUTO ON This command enables a delay on Channel A.
TRIG:SEQ2:DEL:AUTO? This command queries the settling-time
delay of Channel B.