PSTatistic Subsystem 9
Keysight N8262A Programming Guide 337
PSTatistic[1]|2:CCDF:TABle?
This command is used to return the data in CCDF table, average input power,
probability at the average input power, peak to average power ratio and sample
count.
NOTE
This command will return 10 scalar results in the following order:
1 Average input power (in dBm)
2 Probability at the average input power (in %)
3 Power level (power to average power ratio) that has 10 % of the power (in dB)
4 Power level (power to average power ratio) that has 1 % of the power (in dB)
5 Power level (power to average power ratio) that has 0.1 % of the power
(in dB)
6 Power level (power to average power ratio) that has 0.01 % of the power
(in dB)
7 Power level (power to average power ratio) that has 0.001 % of the power
(in dB)
8 Power level (power to average power ratio) that has 0.0001 % of the power
(in dB)
9 Peak to average power ratio (in dB)
10 Sample count
NOTE
This command is only applicable when P-Series sensors are present and the
following conditions are met:
– Free run acquisition mode is selected
– NORMal or DOUBle measurement speed setting is chosen

Table of Contents

Other manuals for Keysight Technologies P Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies P Series and is the answer not in the manual?

Keysight Technologies P Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelP Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals