Power Meter Remote Operation 1
Keysight N8262A Programming Guide 79
Editing Frequency Dependent Offset Tables
It is not possible to create any additional frequency dependent offset tables.
However, the 10 existing tables can be edited using the MEMory subsystem. To do
this:
1 Select one of the existing tables using:
MEMory:TABle:SELect <string>
For information on naming frequency dependent offset tables see “Naming
Frequency Dependent Offset Tables” on page 81. For information on the
current names which you can select refer to “Listing the Frequency Dependent
Offset Table Names” on page 80.
2 Enter the frequency data using:
MEMory:TABle:FREQuency <numeric_value> {,<numeric_value>}
3 Enter the offset factors as shown in the table below using:
MEMory:TABle:GAIN <numeric_value> {,<numeric_value>}
4 If required, rename the frequency dependent offset table using:
MEMory:TABLe:MOVE <string>,<string>. The first <string> parameter
identifies the existing table name, and the second identifies the new table
name.
Frequency Offset
Frequency 1 Offset 1
Frequency 2 Offset 2
""
Frequency n Offset n

Table of Contents

Other manuals for Keysight Technologies P Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies P Series and is the answer not in the manual?

Keysight Technologies P Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelP Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals