PSTatistic Subsystem 9
Keysight N8262A Programming Guide 329
PSTatistic[1]|2:CCDF:DATa?
This command is used to return 501 probability values in % at different power
level within certain range, starts from 0 dB till the predefined maximum power
level.
Syntax
Example
NOTE
The maximum power level can be set by using this command:
PSTatistic[1]|2:CCDF:DATa:MAX <numeric_value>
By default, the maximum value is 50 dB.
The power interval between each reading (probability value) is determined by
the defined maximum power level divided by 500.
NOTE
This command is only applicable when P-Series sensors are present and the
following conditions are met:
– Free run acquisition mode is selected
– NORMal or DOUBle measurement speed setting is chosen
PST
2
:CCDF
:DAT
1
?
PST1:CCDF:DAT? This command returns 501 probability values
in % at different power levels within certain
range (from 0 dB to maximum power level
defined).

Table of Contents

Other manuals for Keysight Technologies P Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies P Series and is the answer not in the manual?

Keysight Technologies P Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelP Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals