DISPlay Subsystem 5
Keysight N8262A Programming Guide 257
Query
DISPlay:[WINDow[1]|2]:FORMat?
The query returns the current format of the selected window.
Query Example
Error Messages
If the command is set to TRACe and the selected channel from which TRACe is
taken has no sensor connected or has on a sensor other than a P-Series power
sensor connected, error –241, “Hardware missing” occurs.
If the command is set to TRACe and the selected channel has a P-Series power
sensor connected in AVERage measurement mode, the error –221, “Settings
conflict” occurs.
DISP:FORM? This command queries the current format of