TRIGger Subsystem 14
Keysight N8262A Programming Guide 575
INITiate:CONTinuous:SEQuence[1]|2 <boolean>
This command sets the power meter for either a single trigger cycle or continuous
trigger cycles. A trigger cycle means that the power meter exits the wait for
trigger state and starts a measurement. When entering local mode,
INITiate:CONTinuous is set to ON.
If INITiate:CONTinuous:SEQuence[1|2] <boolean> is set to:
– OFF, the trigger system remains in the idle state until it is set to ON, or
INITiate:IMMediate is received. Once this trigger cycle is complete the
trigger system returns to the idle state.
– ON, the trigger system is initiated and exits the idle state. On completion of
each trigger cycle, the trigger system immediately commences another trigger
cycle without entering the idle state.
Syntax
Example
NOTE
This command performs the same functions as INITiate[1]|2:CONTinuous
<boolean>.
INIT:CONT:SEQ2 ON This command places Channel B in a wait for
trigger state.
INIT
:CONT
0|OFF
1|ON
Space
?
:SEQ
1
2

Table of Contents

Other manuals for Keysight Technologies P Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies P Series and is the answer not in the manual?

Keysight Technologies P Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelP Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals