16 SERVice Subsystem
634 Keysight N8262A Programming Guide
SERVice:SENSor[1]|2:FREQuency:MINimum?
This query returns the minimum frequency that can be measured by the currently
connected sensor. It is applicable to E-Series sensors only. Minimum frequency
information is stored in the sensor’s EEPROM.
Syntax
Example
Error Messages
If no sensor is connected, error –241, “Hardware missing” occurs.
If a a sensor other than an E-Series power sensor is connected, error –241
“Hardware missing” occurs.
If the E-Series sensor currently connected does not contain the necessary
information in EEPROM, error –241 “Hardware missing” occurs.
SERV:SENS1:FREQ:MIN? This query returns the minimum frequency
that can be measured by the E-Series sensor
currently connected to Channel A.