9 PSTatistic Subsystem
336 Keysight N8262A Programming Guide
Example
Error Messages
– If no power sensor is connected, error –241 "Hardware missing" occurs.
– If sensor/sensors connected are not P-Series sensors, error –241 "Hardware
missing" occurs.
– If measurement speed setting is FAST, error –221 "Settings conflict" occurs.
– If the acquisition mode is in continuous triggering or triggering source is set to
either INT1, INT2, or EXT, error –221 "Setting conflict" occurs.
– If the parameter specified is less than 0.0 or more than 50.0, error –220
"Parameter error" occurs.
– If no parameter is specified, error –109 "Missing parameter" occurs.
PST1:CCDF:PROB? 50 This command queries the probability at the