14 TRIGger Subsystem
578 Keysight N8262A Programming Guide
INITiate[:IMMediate]:SEQuence[1]|2
This command sets the power meter in the wait for trigger state. When a trigger is
received, the measurement is taken and the result placed in the power meter
memory. If TRIGger:SOURce is set to IMMediate the measurement begins as soon
as INITiate:IMMediate is executed.
Use FETCh? to transfer a measurement from memory to the output buffer. Refer to
“FETCh[1]|2|3|4 Queries” on page 145 for further information.
Syntax
Example
Error Messages
If the power meter is not in the “idle” state or INITiate:CONTinuous is ON, error
–213, “INIT ignored” occurs.
NOTE
This command performs the same function as INITiate[1]|2:[IMMediate].
INIT:IMM:SEQ1 This command places Channel A in the
wait for trigger state.
INIT :IMM
:SEQ
1
2