EasyManuals Logo

Keysight Technologies P Series User Manual

Keysight Technologies P Series
673 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #453 background imageLoading...
Page #453 background image
SYSTem Subsystem 12
Keysight N8262A Programming Guide 453
SYSTem:COMMunicate:LAN:ADDRess <character_data>
This command sets the LAN (IP) address of the power meter.
Syntax
Parameters
Example
Query
SYSTem:COMMunicate:LAN:ADDRess?
The query returns the current setting of the LAN address.
Query Example
SYST
:COMM
:LAN
:ADDR
Space
?
character_data
Item Description Range of Values
character_data Numeric character values for the address. Up to 15
characters, formatted as follows:
A.B.C.D where A, B, C, D = 0 to 225
0 to 225 (no embedded spaces)
SYST:COMM:LAN:ADDR
‘130.015.156.255’
This command sets the LAN IP address to
130.015.156.255.
SYST:COMM:LAN:ADDR? This command queries the setting of the LAN
IP address.