5 DISPlay Subsystem
256 Keysight N8262A Programming Guide
Parameters
Example
Reset Condition
On reset, the defaults for the upper and lower windows are DIGital.
Item Description/Default Range of Values
character_data Sets the window format:
– DIGital: sets the window display to digital. This setting is
the same as SNUMeric.
– ANALog: sets the window display to analog using the
currently SELected measurement.
– SNUMeric: sets the window display to single numeric. The
currently SELected measurement is displayed. This setting is
the same as DIGital.
– DNUMeric: sets the window display to dual numeric.
– TRACe: trace display using the currently SELected
measurement. Used to determine the channel from which
the trace is taken.
– CTABle: sets the window display to CCDF table.
DIGital
ANALog
SNUMeric
DNUMeric
TRACe
CTABle
DISP:WIND2:FORM DIG This command sets the lower window to a
digital display.