SERVice:CALibrator:ADJ:FINE <numeric_value>
This command adjust the 1 mW calibrator output in fine scale.
Syntax
Parameters
Query
:CAL
:FINE
numeric_value
DEF
?
SERV
:ADJ
space
Item Description/Default Range of Values
Numeric_value Adjust the 1 mW Power Reference Level Increment Fine by 1. 0 to 1023 (Unsigned Int 16)
SERV:CAL:ADJ:FINE? The query returns the Reference Calibrator
power level in unsigned Int 16.