12 SYSTem Subsystem
512 Keysight N8262A Programming Guide
Table 12-29 802.11b/g: Power meter presets
Command Setting Comments
Frequency
[SENS[1]]|SENS2:FREQ[:CW|:FIX] +2.400 GHz Frequency setting
Sensor measurement mode
[SENS[1]]|SENS2:DET:FUNC NORM Measurement mode
Sensor video bandwidth setup
[SENS[1]]|SENS2:BAND|BWID:VID N1921/2A: HIGH Sensor video bandwidth
Gate Setup
[SENS[1]]|SENS2:SWE[1]|2|3|4:OFF:TIME Gate 1: 0
Gates 2 - 4: 0
Delay between trigger point and time
gated period.
[SENS[1]]|SENS2:SWE[1]|2|3|4:TIME Gate 1: 100 µs
Gates 2 - 4: 0
Length of time gated period for time gated
measurements.
Trigger setup
TRIG[:SEQ[1]|2]:SOUR INT1 Trigger source set up and acquisition
mode continuous triggering
INIT:CONT ON
TRIG[:SEQ]:LEV:AUTO ON Enable automatic setting of the trigger
level
TRIG[:SEQ]:LEV AUTO Power level
TRIG[:SEQ]:SLOP POS Trigger event recognized on the rising
edge of a signal
TRIG[:SEQ]:DEL 0 s Delay between recognition of trigger event
and start of a measurement
TRIG[:SEQ]:HOLD MIN Trigger holdoff
Range
[a]
[SENS[1]]|SENS2:POW:AC:RANG:AUTO OFF Auto range off
[SENS[1]]|SENS2:POW:AC:RANG UPPER Range set to upper
Step detection
[SENSe[1]]|SENS2:AVER:SDET 0 Step detection disabled