10 SENSe Subsystem
376 Keysight N8262A Programming Guide
Reset Condition
On reset, the duty cycle is set to 1 % (DEF).
Query
[SENSe[1]]|SENSe2:CORRection:DCYCle|GAIN3[:INPut][:MAGNitude]?
[MIN|MAX]
The query returns the current setting of the duty cycle or the values associated
with MIN and MAX.
Query Example
Error Messages
– If a duty cycle value is entered using
[SENSe[1]]|SENSe2:CORRection:DCYCle|GAIN3 while
[SENSe[1]]|SENSe2:SPEed is set to 200, the error –221, “Settings Conflict”
occurs. However, the duty cycle value is set but the
[SENSe[1]]|SENSe2:CORRection:DCYCle|GAIN3:STATe command is not
automatically set ON.
– If this command is used when an P-Series or E-Series Power Sensor is
connected, the error –310, “System error;Dty Cyc may impair accuracy with
ECP sensor” occurs. If you are using a dual channel power meter the error
message specifies the channel.
CORR:GAIN3? This command queries the current setting of
the duty cycle for Channel A.