SENSe Subsystem 10
Keysight N8262A Programming Guide 377
[SENSe[1]]|SENSe2:CORRection:DCYCle|GAIN3:STATe <boolean>
This command is used to enable and disable the pulse power measurement
feature.
The [SENSe[1]]|SENSe2:CORRection:DCYCle|GAIN3 command should be used to
enter the duty cycle of the signal you want to measure.
Syntax
Example
Reset Condition
On reset, the pulse power measurement feature is disabled.
Query
[SENSe[1]]|SENSe2:CORRection:DCYCle|GAIN3:STATe?
The query enters a 1 or 0 into the output buffer indicating the status of the pulse
power measurement feature.
– 1 is returned when the pulse power measurement feature is enabled
– 0 is returned when the pulse power measurement feature is disabled
CORR:DCYC:STAT 1 This command enables the pulse
measurement feature on Channel A.
SENS
1
:DCYC
CORR
SENS2
:
:STAT
0|OFF
1|ON
Space
?
:GAIN3

Table of Contents