EasyManuals Logo

Keysight Technologies P Series User Manual

Keysight Technologies P Series
673 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #584 background imageLoading...
Page #584 background image
14 TRIGger Subsystem
584 Keysight N8262A Programming Guide
Parameters
Example
Reset Condition
On reset, the trigger source is set to IMMediate.
Item Description/Default Range of Values
source – Available trigger sources:
– BUS: the trigger source is the group execute trigger <GET> bus command, a *TRG
common command or the TRIGGER:IMMediate SCPI command.
– EXTernal: the trigger source is the trigger input in the front panel.
– HOLD: triggering is suspended. The only way to trigger the power meter is to use
TRIGger:IMMediate.
– IMMediate: the trigger system is always true. If INITiate:CONTinuous is ON
the power meter is continually triggering free (free run mode). If an
INITiate:IMMediate command is sent a measurement is triggered then the
power meter returns to the idle state.
– INTernal: either INT1 (Channel A) or INT2 (Channel B).
BUS
EXTernal
HOLDIMMediate
INTernal[[1]|2]
NOTE
The trigger source is set to IMMediate on instrument power-up and when
entering local mode.
The MEASure and CONFigure commands automatically set the trigger source to
IMMediate.
The READ? or MEASure commands should not be used if the trigger source is set
to BUS or HOLD.
TRIG:SOUR IMM This command configures Channel A for
immediate triggering.

Table of Contents

Other manuals for Keysight Technologies P Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies P Series and is the answer not in the manual?

Keysight Technologies P Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelP Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals