MEASurement Commands 2
Keysight N8262A Programming Guide 181
Example
resolution A numeric value for the resolution. If unspecified the current
resolution setting is used.
1 to 4
[b]
1.0, 0.1, 0.01, 0.001
DEF
[a]
source list This channel list specifies the channels used to calculate the
difference.
If unspecified and the current window setup is a difference
measurement then this difference setup is used, otherwise it
defaults to Channel A-B.
(@1),(@2)
(@2),(@1)
(@1),(@1)
(@2),(@2)
[a] The mnemonic DEF means DEFault. This is not equivalent to the DEFault parameter used in the command sub-systems. The parameters must
be entered in the specified order. If parameters are omitted, they default from the right. The parameter DEFault is used as a place holder.
Specifying DEF leaves the parameter value unchanged.
[b] When the measurement result is linear this parameter represents the number of significant digits. When the measurement result is
logarithmic 1 to 4 represents of 1, 0.1, 0.01 and 0.001 respectively.
MEAS2:RAT? DEF,1,(@1),(@2) This command queries the lower window/
upper measurement ratio measurement of
Channel A over Channel B, using the current
sensor range and a resolution of 1 on both
channels.
Item Description/Default Range of Values