SYSTem Subsystem 12
Keysight N8262A Programming Guide 457
SYSTem:COMMunicate:LAN:HNAMe <character_data>
This command sets the hostname for the power meter.
The factory default setting of hostname is in this format:
K - + product number + - + suffix 5 digits of serial number
For example, K-N8262A-00204
Syntax
Parameters
Example
Query
SYSTem:COMMunicate:LAN:HNAMe?
The query returns the current setting of the LAN hostname.
Query Example
SYST
:COMM
:LAN
:HNAM
Space
?
character_data
Item Description Range of Values
character_data Character values of up to 15 characters Maximum of 15 characters
SYST:COMM:LAN:HNAM
‘PowerMeter1’
This command sets the hostname to
PowerMeter 1.
SYST:COMM:LAN:HNAM? This command queries the setting of the
hostname.